The machine
The machine

 

The GENERA ultrasound system is an innovative method of treating cutaneous ulcer lesions. Rapid radical debridement of the ulcer is obtained through the development of cavitation energy directly in contact with the lesion, removing even the most tenaciously adhered fibrin and necrotic tissue.
Due to the minimal transmission of mechanical effects and heat to the tissue the cavitation cleansing–removal procedure is painless if compared to the use of conventional debridement techniques.

 

The GENERA ultrasound system also achieves a lowering of the bacterial load of the infected ulcer; this effect is the consequence of the removal of adhering infected material and the direct antibacterial action on the bacterial membrane being subjected to permanent damage due to its implosion, provoked by the ultrasonic cavitation. Genera treatment does not interfere with other associated therapeutic measures.
  • RADICAL DEBRIDEMENT OF THE ULCER
  • HANDLING OF THE ULCERATIVE LAYER FOR LACK OF TISSUE DAMAGES
  • LOWERING OF PAIN FOR THE PATIENT IF COMPARED TO THE USE OF CONVENTIONAL DEBRIDEMENT TECHNIQUES
  • LOWERING OF THE BACTERIAL LOAD WITH A STERILIZING EFFECT ON THE INFECTED ULCER
  • ACCELERATION OF THE HEALING PROCESS
  • SPECIFIC METHODS OF TREATING ULCER LESIONS, BY USING 7 DIFFERENT KINDS OF TIPS



The Genera system

This system comprises an ultrasound generator that operates continuously, capable of activating a titanium tip assembled on a piezoelectric transducer and vibrating at a frequency of 26 KHz.

 

With regard to the cleansing of a skin ulcer, there are five types of differently shaped tips available, each capable of producing a specific effect on the lesion to be treated.



 

Tips Codes 610 – 611 and 612 mainly generate an action aimed at detachment of the biological material (fibrinous-necrotic) adhering to the bottom of the ulcerative lesion and at reduction of the bacterial load of infected ulcers. Tip Code 654 performs an abrading action on the fibrinous material in the presence of irregularities of the bottom of the ulcerative lesion with absence of an actual detachment plane. The other 3 tips (Codes 613 – 682 – 666) allow cutting both tissues and small bone structures protruding from the bottom of the ulcerative lesion, which often have a negative effect on ulcer healing. The latter type of tip is used mainly to treat ulcerative lesions on the toes or related to forefoot amputations.

Each of the tips has the distal active part which, under ultrasound activation, is subjected to 26,000 excursions per second (26 KHz) with a transverse vibratory displacement of peak-to-peak amplitude between 15 – 30 µm. The active distal end of the tips is irrigated continuously with the aid of a peristaltic pump.

 

Application principles

The application principles of the Genera system in the treatment of ulcerative lesions depend on the development of mechanical-vibratory energy, cavitational energy and caloric energy. The mechanical-vibratory and cavitational energy generated by the tips allows detaching the fibrinous and necrotic material from the bottom of the ulcerative lesion, with its consequent removal and simultaneous bacterial killing effect. The use of caloric energy obtainable with ultrasonic activation of the tips further increases the system's efficacy in reducing the bacterial load in the presence of infected ulcers.



The sistem genera UDD

 

Rapid and radical ulcer debridement is obtained through the development of cavitational energy directly in contact with the lesion, removing even the most tenaciously adherent fibrinous and necrotic tissues.

The GENERA ultrasound system isan innovative method for treating ulcerative skin lesions.

Owing to the minimal transmission of the mechanical and caloric effects to the tissues, cavitation makes the cleaning-removal procedure less painful than traditional debridement techniques.

The GENERA ultrasound system also produces a decline in bacterial load in infected ulcers. This result stems from the removal of the infected material adhering to the ulcer and from a direct antibacterial action caused by damage on the bacterial membrane, which undergoes implosion following the cavitational effect.

Treatment with Genera does not interfere with the other associated therapeutic measures.

To find out more...

 
Technical

 


GENERA

UDD

ULTRASOUND GENERATOR

Mains power supply 230 VAC ± 10% / 50/60 Hz
Maximum power consumption 140 VA
Ultrasound power 90 Watt
Channels One
Irrigation liquid delivery rate 80 cc/min
Protection fuse 2 x 1.6 A Fast, 250 V
Dimensions
Height 145 mm
Width 300 mm
Depth 330 mm
Weight 4,5 kg

 

HANDPIECE AND TIPS

Technology Piezoelectric
Ultrasound frequency 26 kHz ± 10%
Vibration amplitude 30 μm (peak to peak)
Maximum acceleration peak 48.000 g
Handpiece length 155 mm
Handpiece diameter 23,5 mm
Handpiece sterilization by Ethylene Oxide – Gas Plasma - Autoclave rubber cycle 121° C
Insert tips sterilization by Ethylene Oxide – Gas Plasma – Autoclave 121° C or 134° C


Classification in compliance with standard EN 60601-1: Class I, CF type. Medical device Class II B.

 

 


Italia Medica S.r.l. via Vincenzo Monti 77 - 20149 Milano - Tel. +39.02.33603560 / +39.02.33603577 - P.IVA 10521520154 - c.f. 10110630158